| Item |
Description |
| Titration method |
Volumetric Karl Fischer titration |
| Titration control method |
Intermittent titrant addition while approaching the end point. |
| End point detection |
AC polarization potential difference detection. |
| Buret |
Number of burets: 1, Capacity: 10 mL
Accuracy: within +/-0.15% (for full discharge, n=5)
Reproducibility: +/-0.01 mL (for full discharge, standard deviation for n=5) |
| Measuring range |
0.1 mgH2O - 500 mgH2O (10 ppm - 100%) |
| Minimum volume of addition |
0.005 mL |
| Titration cell capacity |
150 mL
(Minimum volume of consumption: About 30 mL) |
| Display |
16-Digit x 2-Line LCD with backlighting |
| Display unit |
mL (min. display: 0.001mL), mgH2, %, ppm |
| Calculation function |
Concentration calculation, Factor standardization, Statistics calculation, recalculation |
| File function |
4 files for titration parameters |
| Memory data capacity |
20 data |
| Printers |
Optional impact printer; PR-302B(CBM-910-40PF) |
| External input / output |
RS-232C; 2 ports
Port 1 (BALANCE):Data input from balance or data input.
Port 2 (HOST):Command instructions from host computer. |
| Power supply / Power consumption |
AC 100 - 240 V, 50/60 Hz, 100 VA |
| Dimensions / Weight |
310(W) x 270(D) x 300(H)mm
Approx. 6 kg |