Autoclave Compatible Micropipettes

Back

Nichipet EXII

Nichiryo Nichipet EXII is an autoclavable digital micropipette.

Whatsapp

Overview

Features

1. Glass-made counter window

Volume display made of durable glass material protects the damage from UV irradiation and keep the pipette safer in clean bench even over a long time.

2. Lock mechanism

Designed with a lock mechanism,it is easier to adjust volume smoothly.

3. Body

Materials used for pipette body have high tolerance to solvents.
Non-grease designed. No need to worry about continuous autoclaving.

4. Ejector pipe

The ejector pipe fixed by a screw is easier to remove.

5. Plunger

Non corrosive ceramic used in the plunger. (200uL-10000uL)

6. Resin-made ejector pipe

Resin material of ejector pipe could enhance high tolerance to solvents.

7. Hyper Blower system

Hyper blower system improves dispensing efficiency, especially on 2μL and on 10μL dispensing.

 

Best Selling Model of Nichipet Series

  • Fully autoclavable. ( 121°C for 20 minutes )
  • Enhanced UV resistance.
  • Hyper Blower System, longer second - push stroke, improves dispensing efficiency.(2μL, 10μL)
  • Easy digital volume setting.
  • Easy and convenient single hand locking lever.
  • Eight sizes to cover a wide volume range. ( 0.1μL - 10000μL )
  • PTFE ( Fluorine resin ) is used in the airtight chamber of the instrument, so the instrument keeps its air tightness and precise reproducibility for long hours.
  • Ceramic plungers are used for volume models from 200μL to 10000μL.
  • Filtered Nozzles are used for volume models from 1000μL to 10000μL.
  • Tip ejector allows tip ejections without touching the tip.
  • Tip ejector is made of plastic to avoid damaging glass tubes.

 

UV ResistanceAutoclavable

 

Specifications

Cat. No Volume Setting Volume range(μL) Increments(μL) Usable tips
00-NPX2-2 Digital Setting 0.1?2 0.002 BMT2-UT/UTWR, FUT/FUTRB
00-NPX2-10 0.5?10 0.01 BMT2-SSW/SSWR/SSWRB, 
FSS/FSSRB
00-NPX2-20 2?20 0.02 BMT2-SG/SGR, SE/SER, FSG/FSGRB
00-NPX2-100 10?100 0.1 BMT2-SG/SGR, SE/SER, FSG/FSGRB*. FSE/FSERB
00-NPX2-200 20?200 0.2 BMT2-SG/SGR, SE/SER, FSE/FSERB, FAG/FAGRB
00-NPX2-1000 100?1000 1 BMT2-LG/LGR, LE/LER, FLG/FLGRB, FLE/FLERB
00-NPX2-5000 1000?5000 10 BMT2-X/XR
00-NPX2-10000 1000?10000 10 BMT2-Z/ZR

 

AC & CV

Cat. No Volume(μL) Accuracy(%) Precision(%)
00-NPX2-2 0.2 ±12.0* ≤6.0*
1 ±5.0 ≤2.5
2 ±3.0 ≤1.0
00-NPX2-10 1 ±4.0 ≤3.0
5 ±1.0 ≤1.0
10 ±1.0 ≤0.5
00-NPX2-20 2 ±5.0 ≤3.0
10 ±1.0 ≤1.0
20 ±1.0 ≤0.4
00-NPX2-100 10 ±2.0 ≤1.0
50 ±1.0 ≤0.3
100 ±0.8 ≤0.3

 

 

Cat. No Volume(μL) Accuracy(%) Precision(%)
00-NPX2-200 20 ±1.0 ≤0.5
100 ±0.8 ≤0.3
200 ±0.8 ≤0.2
00-NPX2-1000 100 ±1.0 ≤0.5
500 ±0.8 ≤0.3
1000 ±0.7 ≤0.2
00-NPX2-5000 1000 ±1.0 ≤0.3
2500 ±0.8 ≤0.3
5000 ±0.6 ≤0.2
00-NPX2-10000 1000 ±2.0 ≤0.4
5000 ±0.8 ≤0.3
10000 ±0.4 ≤0.2
  • *The AC and CV values below 0.2μL depend much on the operator’s skill and the environment under which the pipette is used.
  • *The AC and CV values are the values by use of the disposable tips described in the catalogue.
  • *The specification and price are subjected to change without notice due to modifications for quality improvement.
  • *The Nichipet EX Plus II is made of parts resistant to organic solvents which makes it more durable than previous pipettes in environments with organic solvents. However, it does not guarantee accuracy of dispensing organic solvents.
  • *As the body of the Nichipet EX Plus II is not resistant to organic solvents, make sure it does not come in contact with the organic solvents.

 

Stands

00-MLT-STD2(ABS)

Cat. No Storing capacity
00-MLT-STD2(ABS) 6pcs

 

 

00-BM-STD(Acrylic resin)

Cat. No Storing capacity
00-BM-STD(Acrylic resin) 6pcs

 

 

S-PAC

S-PAC

Cat. No Content
00-SP-A2
Pipette : NPX2-10(0.5~10μL)
NPX2-100(10~100μL)
NPX2-1000(100~1000μL)
Tip : BMT2-SSR(0.5~10μL)
BMT2-SGR(2~200μL)
BMT2-LGR(100~1000μL)

Tube Opener:NTO-001 (3pcs)

00-SP-BN2
Pipette : NPX2-20(2~20μL)
NPX2-200(20~200μL)
NPX2-1000(100~1000μL)
Tip : BMT2-SGR(2~200μL)×2
BMT2-LGR(100~1000μL)

Tube Opener:NTO-001 (3pcs)

00-SP-H2
Pipette : NPX2-2(0.1~2μL)
NPX2-10(0.5~10μL)
NPX2-100(10~100μL)
Tip : BMT2-UTWR(0.1~10μL)×2
BMT2-SSR(0.5~10μL)
BMT2-SGR(2~200μL)

Tube Opener:NTO-001 (3pcs)

Each package contains 3 Nichipet EXIIs, 3 sample racked tips and 3 tube openers.