Four-Channel Free Radical Analyzer

Back

Four-Channel Free Radical Analyzer - TBR4100

Fast, reliable, real-time detection – measure redox-reactive species

Whatsapp

Overview

Features:

  • Real-time detection using electrochemical microsensors
  • Integrated system includes one temperature sensor, your choice of two additional sensors and a start-up kit
  • Current measurement range from 300 fA to 10 µA (four ranges) permits wide dynamic range for detection
  • Wide bandwidth allows recording of fast events
  • Measure carbon monoxide from 10 nM to 10 µM
  • Measure nitric oxide from < 0.3 nM to 100 µM
  • Measure hydrogen peroxide < 10 nM to 100 mM
  • Measure hydrogen sulfide
  • Measure glucose
  • Measure oxygen from 0.1% to 100%
  • Isolated architecture allows Lab-Trax interface to simultaneously measure free radical and independent analog data (for example, ECG, BP, etc.) on any channel
  • Four channel free radical detection

Specifications:

Power 100 ~ 240 VAC, 50-60 Hz,
Operating Temperature (ambient) 0 - 50°C (32 - 122°F)
Operating Humidity (ambient) 15 - 70% RH non-condensing
Warm up Time < 5 min.
Dimensions 135 X 419 X 217 mm (5.25" X 16.5" X 8.16")
Weight 1.35 kg (3 lb.)
Display Functions 18 mm (0.7") LCD readout, 4.5 digit Polarization Voltage (mV) Current input (nA, µA)
Controls Power (on/off)
Current Input Range 
Polarization Voltage
Analog Output Range ±10 V (continuous)
Analog Output Impedance 10 KΩ
Channel to Channel Isolation >10 GΩ
Channel to Output Isolation >10 GΩ
Power Supply to AC Line Isolation >100 MΩ
Analog Output Drift < 10 pA/hr.
Temperature Input: Number of Channels 1
Temperature Input: Sensing Element Platinum RTD, 1000 Ω
Temperature Input: Range 0-100°C
Temperature Input: Accuracy ± 1°C
Temperature Input: Resolution 0.1°C
Temperature Input: Analog Output 31.25 mV/°C (continuous)
Amperometric Input: Number of Amperometric Channels 4
Amperometric Input: Signal Bandwidth 0-3 Hz
Amperometric Input: Polarization Voltage (selectable via rotary switch) Nitric Oxide 865 mV
Amperometric Input: Polarization Voltage (selectable via rotary switch) Hydrogen Sulfide 150 mV
Amperometric Input: Polarization Voltage (selectable via rotary switch) Hydrogen Peroxide 450 mV
Amperometric Input: Polarization Voltage (selectable via rotary switch) Glucose 600 mV
Amperometric Input: Polarization Voltage (selectable via rotary switch) Oxygen 700 mV
Amperometric Input: Polarization Voltage (selectable via rotary switch) ADJ (user adjustable) ± 2500 mV
Polarization Voltage Accuracy ± 5 mV
Polarization Voltage Display Resolution ± 1mV
Current measurement Performance: 
Range  Analog Output Noise @ 3 Hz* Noise @ 0.3 Hz*
±10 Na 1 mV / 1 pA < 1 pA < 0.3 pA
± 100 nA 1 mV / 10pA < 7 pA < 3 pA
± 1 µA 1 mV / 100pA < 70 pA < 30 pA
±10 µA 1 mV / 1µA < 700 pA < 300 pA
Notes: *Instrument performance is measured as the (max-min) over 20 seconds period with open input. Typical values are given at 3 Hz and 0.3 Hz bandwidth.
Typical sensor performance with TBR4100: ISO-NOPF100 noise 0.2 nM NO (< 2pA **)
Notes: **Sensor noise is measured as the (max-min) over a 20 seconds period with the sensor immersed in 0.1 M CuCl2 solution.